A-A-54092
Shall have minimum draws of 8.5 ml and 11.5 ml for the specification designated 9.5 ml 13 ml aspiration capacity tubes respectively.
Shall comply with the applicable requirements of NCCLS (National Committee for Clinical Standards) Approved Standard ASH-1 (Standard for Evacuated Tubes for Blood Specimen Collection.
Shall be suitable for use with a standard double-ended blood-collecting needle/holder, described in the ASH-1 standard, that contains a bevel edge needle in its open end. When the stopper ends of the tubes are inserted into the needle/holder to the depth even with the needle/holder’s reference mark, the needle point shall be embedded in the stoppers without puncturing the stoppers.
Workmanship. The blood collecting tubes shall be free from defects which detract from their appearance or impair their serviceability.
Unit. Package (PG). One package containing one thousand tubes, as specified, constitutes one unit.
Quality Assurance Provisions.
Responsibility for inspection. Unless otherwise specified in the contract or purchase order, the contractor is responsible for the performance of all inspection requirements as specified herein. Except as otherwise specified in the contact or purchase order, the contractor may use his own or any facilities suitable for the performance of the inspection requirements specified herein, unless disapproved by the Government. The Government reserves the right to perform any of the inspections set forth in the specification where such inspections are deemed necessary to assure supplies and services conform to prescribed requirements.
Records. Records of examinations and tests performed by or for the contractor shall be maintained by the contractor and made available to the Government upon the Government’s request, at any time, or from time to time, during the performance of the contract and for a period of three years after delivery of the supplies to which such records relate.
Inspection. Inspection, as used herein, is defined as both examination (such as visual or auditory investigation without the use of special laboratory appliances or procedures) and testing (determination by technical means of physical and chemical properties) of the item.